Model 7i Digital Force / Torque Indicator User’s Guide
34
Press ENTER.
11. At the completion of the tension calibration, the display appears as follows:
To save the calibration information, select “Save & exit”. To exit without saving the data select
“Exit without saving”.
12. Any errors are reported by the following screens:
Displayed at the start of calibration if a disallowed unit is selected.
Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.
CALIBRATION
Load not stable.
Please try again.
CALIBRATION
Units must be gF.
Please try again
Press ENTER.
CALIBRATION
COMPRESSION
Load too low.
Please try again.
CALIBRATION
COMPLETE
Save & exit
Exit w/o saving
CALIBRATION
COMPRESSION COMPLETE
Reverse direction
for tension.
Attach necessary
weight fixtures,
then
p
ress ENTER.