Cooper-instruments-systems Series M4 Digital Force Gage Manual do Utilizador

Consulte online ou descarregue Manual do Utilizador para Instrumentos de medida Cooper-instruments-systems Series M4 Digital Force Gage. Cooper Instruments & Systems Series M4 Digital Force Gage User Manual Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - Series 4 Digital Force Gages

CF 182 32-1117 1010 Users Guide Series 4 Digital Force Gages www.cooperinstruments.com

Página 2

CF 182 7 32-1117 1010 or alarm indication in process control applications. Two limits, high and low, are specified and s

Página 3 - Table of Contents

CF 182 8 32-1117 1010 8.1 View Data All the saved data points may be viewed. The record number is displayed, along with

Página 4 - Caution!

CF 182 9 32-1117 1010 9.1 Installing the USB driver It is recommended that the USB driver be installed before physicall

Página 5 - 2 POWER

CF 182 10 32-1117 1010 Windows 7 Operating Systems – When the USB device has been plugged into a USB port, the driver wi

Página 6 - 3 MECHANICAL SETUP

CF 182 11 32-1117 1010 Click “Continue Anyway”. 8. The next, and final, screen appears as follows: Click “Finish”. T

Página 7 - 4 HOME SCREEN AND CONTROLS

CF 182 12 32-1117 1010 Baud Rate: 9,600 Data Format: Numeric + units Other communication settings are permanently set t

Página 8 - 4.3 Menu navigation basics

CF 182 13 32-1117 1010 9.6 I/O Connector Pin Diagram (female) Pin No. Description Input / Output 1 Signal Ground

Página 9 - 7.1 General Information

CF 182 14 32-1117 1010 The gauge can be calibrated at up to 10 points in each direction. Enter the number of calibration

Página 10 - 8 DATA MEMORY AND STATISTICS

CF 182 15 32-1117 1010 Optionally exercise the load cell shaft several times (at full scale, if possible), then press E

Página 11 - 9 COMMUNICATIONS

CF 182 16 32-1117 1010 Reverse the orientation of the load cell shaft by rotating the gauge 180 degrees. Press DIRECTI

Página 12

CF 182 32-1117 1010 This page intentionally blank.

Página 13

CF 182 17 32-1117 1010 11 OTHER SETTINGS 11.1 Automatic Shutoff The gauge may be configured to automatically power off f

Página 14 - 9.2 Serial / USB

CF 182 18 32-1117 1010 11.5 Information / Welcome Screen The following screen is displayed at power up and can be access

Página 15 - 9.5 DATA Key Functions

CF 182 19 32-1117 1010 Automatic Shutoff Disabled Minutes (if enabled) 5 Beeps Keys Alerts Set Points Enabled Enab

Página 16 - 10 CALIBRATION

CF 182 20 32-1117 1010 13 WARRANTY REPAIR POLICY Limited Warranty on Products Any Cooper Instruments product which, und

Página 17

CF 182 ii 32-1117 1010 Table of Contents 1 OVERVIEW...

Página 18

CF 182 1 32-1117 1010 Thank you for purchasing a Series 4 digital force gauge, designed for tension and compression forc

Página 19

CF 182 2 32-1117 1010 The following safety checks and procedures should be performed before and during operation: 1. N

Página 20 - 11 OTHER SETTINGS

CF 182 3 32-1117 1010 43 MECHANICAL SETUP 3.1 Loading shaft orientation In order to accommodate a variety of testing req

Página 21 - 12 SPECIFICATIONS

CF 182 4 32-1117 1010 4 HOME SCREEN AND CONTROLS 4.1 Home Screen No. Name Description 1 Measurement direction - indi

Página 22 - 12.4 Dimensions

CF 182 5 32-1117 1010 – the displayed value is greater than the upper force limit – the displayed value is between t

Página 23 - 13 WARRANTY REPAIR POLICY

CF 182 6 32-1117 1010 Refer to the following sections for details about setting up particular functions and parameters.

Comentários a estes Manuais

Sem comentários